Gaeltec News

NEEDLE AND CATHETER TYPES

Several needle and catheter types exist for use of measurement of intracompartmental pressures. Below we discuss each of the types and present the advantages and disadvantages associated with each. 

Simple Needle

Early compartmental measurement techniques employed a simple 18 gauge needle inserted into the compartment to be measured. The needle was then connected to a mercury pressure manometer for pressure recording. Although the needles are readily available, several studies have shown that they yield the least reliable and least reproducible data. Because there is a single entry port for interstitial fluid, pressures are falsely elevated or depressed if the needle becomes plugged with tissue or is inadvertently inserted into tendon or fibrous tissue. 

Wick Catheter

The wick catheter technique was introduced in 1976 to provide an improved method for pressure monitoring. The catheter contains multiple fibers at its tip which are introduced into a compartment with an introducer needle. The fibers act by capillary action to improve rapid equilibration with intracompartmental interstitial pressures, while preventing ball-valve obstruction at the tip. Furthermore, the catheter may be left in place for up to 8 hours for serial recordings. Disadvantages include the possibility of coagulation and occlusion around the tip, as well as the possibility for fibers to break off and remain within the tissues. Furthermore, wick catheters may only be used to measure a single compartment and cannot be reused. 

Slit Catheter

The slit catheter was introduced in 1981. It is a polyethelene catheter with 5 six-millimeter axially cut slits at the tip to provide a greater open surface area of the catheter for equilibration with interstitial fluid. As with the wick catheter, it may be left in place for serial pressure recordings, but is also susceptible to clotting. Additionally, it may only be used for measurement of a single compartment.

Side-Ported Needle






The side-ported needle was introduced in 1988. These needles are inserted into intramuscular compartments perpendicular to the surface, and therefore side ports have low risk of occlusion. This allows for improved accuracy (compared with simple needles) of interstitial pressure measurements along with the additional advantage of being capable of measuring several compartments in the same patient 43 . They have been shown to have similar accuracy to slit catheters. The Stryker Pressure Monitor utilizes this needle type.

Transducer-Tipped Catheters


These are open-ended catheters that have a built-in transducer tips. The advantage is that no saline needs to be injected into the measured compartment (which would have the potential to further increase compartmental pressures). There is no needed external manometer or pressure transducer, and the catheter is directly connected to the display. Catheters are relatively expensive, but may be sterilized and reused. They are available from Gaeltec.com.

Useful information

See our information section for anything from "What a Gaeltec Pressure Transducer is" to "Can I have a Transducer with Double Lumen, Twin Bore, Pig Tail with an Extended Tip?" Click here

Making a Gaeltec Catheter

Skye photo gallery