General PurposeLuer Fitting Pressure Transducers

An end-mounted sensor within an electrically isolated titanium shell for measuring physiological pressures

The transducer's extremely small sensing area and volume displacement allow for excellent frequency response.

This transducer, for measuring physiological pressures externally, has an end-mounted sensor within an electrically isolated titanium shell which is tapered to match standard female luer locking fittings. It is supplied with a captive luer lock nut.

The transducer weighs less than 4gm and can be operated with a wide range of pressure amplifiers.

Sterilisation is by means of cold aqueous solutions of detergicide (e.g. Cidex) or by ethylene oxide gas. Do not use sodium hypochlorite solution.

Technical details:

Sensor:

Metal diaphragm with directly deposited resistive strain gauges

Excitation:

5V AC r.m.s. maximum or 1V DC maximum

Bridge resistance:

1.5kΩ nominal

Sensitivity:

5µV/V/mmHg

Linear pressure range:

0 - 300mmHg

Compensated temperature range:

15 - 40°C

Temperature coefficient of zero:

< 0.05%FS/C°

Temperature coefficient of sensitivity:

< 0.2%/C°

Linearity and hysteresis error:

< ±1%FS BSL

Overpressure:

1200mmHg

Cable:

3 metres of 3mm Ø with silicone rubber jacket

Connector:

Please specify the amplifier type or connector when ordering